Amendment 1:2018 to CAN/CSA-C22.2 NO. 61010-1-12, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements (Adopted IEC 61010- 1:2010) (Tri-national standard, with UL 61010-1 and ANSI/ISA-61010-1 (82.02.01)
This is the harmonized CSA Group, ISA, and ULSE standard for Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements. It is the third edition of CAN/CSA-C22.2 No. 61010-1, the third edition of ANSI/ISA-61010-1 (82.02.01), and the third edition of UL 61010-1. This edition of CAN/CSA-C22.2 No. 61010-1 supersedes the previous edition published as CAN/CSA-C22.2 No. 61010-1 in 2004. This harmonized standard has been jointly revised on November 15, 2024. For this purpose, CSA Group and ULSE are issuing revision pages dated November 15, 2024, and ISA is issuing the changes to the edition dated November 15, 2024. This harmonized standard is based on IEC Publication 61010-1: third edition Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General requirements issued June 2010, as revised by IEC Corrigendum 1 issued May 2011, AMD1 issued December 2016, and Corrigendum 1 to AMD1 issued March 2019. IEC publication 61010-1 is copyrighted by the IEC. This harmonized standard was prepared by CSA Group, the International Society of Automation (ISA), and ULSE. This standard is considered suitable for use for conformity assessment within the stated scope of the standard. This standard was reviewed by the CSA Subcommittee on Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, under the jurisdiction of the CSA Technical Committee on Consumer and Commercial Products and the CSA Strategic Steering Committee on Requirements for Electrical Safety, and has been formally approved by the CSA Technical Committee. This standard has been developed in compliance with the Standards Council of Canada requirements for National Standards of Canada. It has been published as a National Standard of Canada by CSA Group.
and object 1.1 Scope 1.1.1 Equipment included in scope This group safety publication is primarily intended to be used as a product safety standard for the products mentioned in the scope, but shall also be used by technical committees in the preparation of their publications for products similar to those mentioned in the scope of this standard, in accordance with the principles laid down in IEC Guide 104 and ISO/IEC Guide 51. This part of IEC 61010 specifies general safety requirements for the following types of electrical equipment and their accessories, wherever they are intended to be used.
This standard does not apply to equipment within the scope of
This standard applies only to computers, processors, etc. which form part of equipment within the scope of this standard or are designed for use exclusively with the equipment. NOTE Computing devices and similar equipment within the scope of IEC 60950 and conforming to its requirements are considered to be suitable for use with equipment within the scope of this standard. However, some of the requirements of IEC 60950 for resistance to moisture and liquids are less stringent than those in this standard (see 5.4.4 second paragraph)). 1.1.3DV DR Modification of Clause 1.1.3 NOTE to revise as follows: Replace "IEC 60950" with "IEC 60950 or IEC 62368-1" in two places. 1.1.4DV DR Addition of the following referencing the National Electrical Code and the Canadian Electrical Code: This standard applies to equipment
Note 1 This standard is considered not to apply to gas detectors and similar equipment within the scope of ANSI/UL 2034, ANSI/UL 2075, CSA C22.2 No. 205, or ULC-S588. Note 2 Examples of standards providing additional safety requirements for gas or vapor detectors or sensors, which can be used in conjunction with this standard, are ANSI/UL 60079-29-1, ANSI/UL 920001, ANSI/UL 920004, CSA C22.2 No. 60079-29-1, or CSA C22.2 No. 62990-1. Note 3 Clause 1.2.1 specifies the use of Clause 17 to assess conformity for hazards or environments not fully covered in Clauses 6 to 16, including aspects not covered by this document in Clause 1.2.2. Note 4 Local and national codes may have additional requirements based on the end products' intended use. Note 5 This standard is considered not to apply to voltage detectors and similar equipment within the scope of UL 1436, CAN/ULC 61243 series, or CSA C22.2 No 160
The purpose of the requirements of this standard is to ensure that HAZARDS to the OPERATOR and the surrounding area are reduced to a tolerable level. Requirements for protection against particular types of HAZARD are given in Clauses 6 to 13, as follows
This standard does not cover
This standard also specifies methods of verifying that the equipment meets the requirements of this standard, through inspection, TYPE TESTS, ROUTINE TESTS, and RISK assessment.
This standard applies to equipment designed to be safe at least under the following conditions
This standard applies to equipment designed to be safe not only in the environmental conditions specified in 1.4.1, but also in any of the following conditions as RATED by the manufacturer of the equipment