IEC 61207-2 Ed. 2.0 b:2019

Current Revision
Expression of performance of gas analyzers - Part 2: Measuring oxygen in gas utilizing high-temperature electrochemical sensors

Best Price Guarantee Jun 1, 2019 EN/FR 46 Pages

Sub Total (1 Item(s)) $ 0.00
Estimated Shipping $ 0.00
Total (Pre-Tax) $ 0.00
View in Library
or
International Electrotechnical Commission Logo

Expression of performance of gas analyzers - Part 2: Measuring oxygen in gas utilizing high-temperature electrochemical sensors

PUBLISH DATE 2019
PAGES 46

Expression of performance of gas analyzers - Part 2: Measuring oxygen in gas utilizing high-temperature electrochemical sensors.

IEC 61207-2:2019 applies to all aspects of analyzers using high-temperature electrochemical sensors for the measurement of oxygen in gas. It applies to in-situ and extractive analyzers and to analyzers installed indoors and outdoors.

The object of this part is:

  • to specify the terminology and definitions related to the functional performance of gas analyzers, utilizing a high-temperature electrochemical sensor, for the continuous measurement of oxygen concentration in a sample of gas;
  • to unify methods used in making and verifying statements on the functional performance of such analyzers;
  • to specify what tests are performed to determine the functional performance and how such tests are carried out;
  • to provide basic documents to support the application of internationally recognized quality management standards.

This second edition cancels and replaces the first edition published in 1994. This edition constitutes a technical revision.

This edition includes the following significant technical changes with respect to the previous edition:

  • a) all the terms and definitions relating to the document have been updated where appropriate;
  • b) the description of the principle of the galvanic cell has been expanded and clarified;
  • c) new definitions and illustrations have been added for different measurement methods for oxygen using solid electrolytes for galvanic cells;
  • d) new illustrations have been added for existing descriptions for ion pump cells;
  • e) a more detailed description of the effect of the presence of oxidizable gases has been added;
  • f) all references to errors have been replaced by uncertainties and appropriate updated definitions applied.